Olympus High-Flow Insufflation Unit, Model UHI-4. For laparoscopic surgery.
FDA Recall Z-0075-2024 · Posted 2023-10-18
Advertisement
Recalling FirmOlympus Corporation of the Americas
Recall StatusOpen, Classified
Product CodeHIF
Date Initiated2023-09-22
Product Quantity3136 units
Reason for Recall
There have been reports of patients suffering arrhythmias, reported as short cardiac arrests, during surgical procedures where UHI-4s were used. These events may have been due to an over insufflation of the abdominal cavity resulting from use of the UHI-4 during the procedures.
Firm Action
An URGENT: MEDICAL DEVICE CORRECTIVE ACTION notification letter dated 9/22/23 was sent to customers.
Action steps to be taken by the end user:
Our records indicate that your facility has purchased one or more of the Olympus UHI-4. Olympus requires you to take the following actions:
1. Carefully read the content of this Field Corrective Action Letter.
2. Ensure all personnel are completely knowledgeable and thoroughly trained on the content of this letter.
3. Complete the enclosed response form and return it to our recall partner, Sedgwick, via e-mail at [email protected] or by fax at 866-367-6717. For any questions about the acknowledgement form, please call the Sedgwick team at 855-215-4972.
4. If you have further distributed this product, identify your customers, and forward them this Letter. Please appropriately document your notification process and let us know the end-customer feedback accordingly.
Olympus requests that you report complaints, including any injuries associated with over insufflation during the procedure with UHI-4, to Olympus. Please report complaints to Technical Assistance Center (TAC) at 1-800-848-9024, option 1. Adverse events experienced with the use of this product may also be reported to the Food and Drug Administration MedWatch Adverse Event Reporting program either online, by regular mail or by fax.
An URGENT: MEDICAL DEVICE CORRECTIVE ACTION - UPDATE notification letter dated 10/25/23 was sent to customers.
Action steps to be taken by the end user:
Our records indicate that your facility has purchased one or more of the Olympus UHI-4. Olympus requires you to take the following actions:
1. Inspect your inventory and identify any UHI-4 High flow insufflation unit(s). Quarantine and mark your units appropriately by your site to prevent usage. If your facility does not have alternatives or is unable to obtain alternatives, you may choose to use the UHI-4 while using extreme caution and after reviewing the information above
Root Cause
Under Investigation by firm
Distribution Pattern
US Nationwide distribution.