CryoCath Frost Byte CryoSurgical Clamp, Model 60FB1, Manufacturer CryoCath Technologies, Inc., Quebec, Canada. The product consists of a cryosurgical probe (SurgiFrost 7 cm) plus Clamp (FrostByte) packaged for use with Cryosurgical console.

FDA Recall Z-0079-2007 · Posted 2006-10-28

Recalling FirmATS Medical, Inc.
Recall StatusTerminated
Product CodeGEH
Date Initiated2006-05-05
Date Terminated2006-11-02
Product Quantity219 units

Reason for Recall

Incorrect Expiration Date: Three lots of the FrostByte (TM) clamp packaged with SurgiFrost 7 cm cryosurgical probes have incorrect expiration dates. These lots are labeled with a two (2) year expiratation date while the correct date is one (1) year.

Firm Action

Consignee''s were notified by ATS Medical Representatives via visits. Product was either relabeled with correct expiration date or retrieved by ATS Medical representatives.

Root Cause

Other

Distribution Pattern

Nationwide-ATS Medical distirubted devices to hospitals located in the following states IN, KS, MO, TX. There may be other US distributors of this product.