Terminated

Intuitive Surgical, daVinci S Surgical System, Model Number: IS2000, A5.1 P1 and A5.0 P5 software level, Surgical Endoscopic Instrument Control System, Intuitive Surgical, Sunnyvale, CA 94086.

FDA Recall Z-0079-2008 · Posted 2008-02-21

Recall StatusTerminated
Product CodeNAY
Date Initiated2007-03-30
Date Terminated2008-09-05
Product Quantity159

Reason for Recall

System Lock-Up: Software anomalies could cause product failure during use, or on start-up. System transitions to a safe "soft-lock" state.

Firm Action

Intuitive Field Service Engineers installed a software upgrade beginning on 03/30/2007, and left a notification letter that the upgrade had been performed at each site. New software level is A5.1 P2.

Root Cause

Software design

Distribution Pattern

Worldwide Distribution - USA, Belgium, China, France, India, Italy, Japan, Korea, Spain, Sweden, Switzerland, UK and Mexico.