Leveen Coaccess Needle Electrode (3.5/15) and (4.0/15). Disposable, monopolar, electrosurgical devices, intended for use in conjunction with BSC radiofrequency generator for the thermal coagulation necrosis of soft tissues, including partial or complete ablation of nonresectable liver lesions.
FDA Recall Z-0080-2013 · Posted 2012-10-18
Advertisement
Recalling FirmBoston Scientific Corporation
Recall StatusTerminated
Product CodeJOS
Date Initiated2012-09-17
Date Terminated2013-07-17
Product Quantity19,716 units distributed total
Reason for Recall
Leveen Standard and Leveen Coaccess Needle Electrode System, Directions for Use(DFU) in Portuguese, are incorrect. all other languages in the DFU are correct. Consignees in Brazil, Portugal,and Hong-Kong are being notified.
Firm Action
Boston Scientific sent notifications, dated September 17, 2012, to each account in Brazil, Portugal and Hong Kong affected by the recall. The affected accounts were provided the corrected pages of the Portuguese DFU as part of the communication to ensure that those users that rely on the Portuguese DFU have the correct information.
Root Cause
Labeling Change Control
Distribution Pattern
Distributed in Brazil, Portugal and Hong Kong.