Foundation Hip System Bipolar Modular, Size 42mm OD; Cat. #412-01-042; Sterile; assembly required by physician prior to implant; manufactured by Encore Medical, L.P., Austin, TX 78758.

FDA Recall Z-0084-2008 · Posted 2007-11-07

Recalling FirmEncore Medical, Lp
Recall StatusTerminated
Product CodeKWY
Date Initiated2007-08-02
Date Terminated2008-01-09
Product Quantity382 units.

Reason for Recall

Device mislabeled. Hip prosthesis components validated for 5 years were labeled as having a 6 year expiration date.

Firm Action

Firm notified consignees of recall via phone, e-mail and overnight letters beginning on 08/02/07. Consignees asked to return devices for a properly labeled replacement.

Root Cause

Packaging change control

Distribution Pattern

Worldwide, including USA, Greece, Japan, and Saudi Arabia.