Siemens Mammomat Inspiration mammography systems. The Mammomat Inspiration system is intended for mammography exams, screening, diagnosis, and stereotactic biopsies under the supervision of medical professionals. Mammographic images can be interpreted by either hard copy film or soft copy workstation.

FDA Recall Z-0086-2015 · Posted 2014-10-17

Recall StatusTerminated
Product CodeMUE
Date Initiated2014-09-18
Date Terminated2015-02-24
Product Quantity3

Reason for Recall

The stereo biopsy devices for Mammomat Inspiration mammography systems might have integrated a safety switch which causes a failure of the functionality. The pin implemented in the safety switch may not put enough pressure on the safety circuit to prohibit movement. The needle positioning device may move even with the safety switch being set sideward. If this occurs with the needle being already i

Firm Action

A safety advisory notice, dated September 18, 2014, was sent to end users that identified the product, problem, and actions to be taken.

Root Cause

Process control

Distribution Pattern

US Distribution in states of: NY, VA, and MA.