Terminated

Endoskeleton" TCS Interbody System with Titan nanoLOCK" Surface Technology, NanoLock, 6 Deg, Interbody, Medium, 10mm

FDA Recall Z-0086-2022 · Posted None

Recall StatusTerminated
Product CodeOVE
Date Initiated2021-09-02
Date Terminated2023-10-05
Product Quantity9 units

Reason for Recall

Engraving on implants may not match what is listed on the pouch labels.

Firm Action

A customer letter was sent via FedEx to impacted US consignees and delivered on or about September 3, 2021. Customers are asked to locate and remove the impacted product from normal storage locations. Customers are advised to not use product. A Customer Confirmation Form should be completed and returned to [email protected]. Customers may also coordinate return and replacement of impacted product with the Medtronic Sales Representative. This recall notice should be forwarded to those within the customer organization or to any organization where potentially impacted devices have been transferred.

Root Cause

Packaging change control

Distribution Pattern

Distribution in US: CA, MN, MO, NC, NV, NY, OH, OK, OR, PA, SC, SD, WA, WV

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.

Official Source

View FDA Recall →