Draeger Medical Apollo Anesthesia Machine, catalog number 8606500

FDA Recall Z-0087-2009 · Posted 2008-10-08

Recalling FirmDraeger Medical, Inc.
Recall StatusTerminated
Product CodeBSZ
Date Initiated2008-08-22
Date Terminated2009-09-01
Product Quantity214 units

Reason for Recall

Sporadic errors in various device functions, including low readings of the single gas flows for oxygen, nitrous oxide, or air delivery, or mechanical ventilation failure. Alarms and error messages function as designed.

Firm Action

Draeger Medical sent a Medical Device Recall letter to their customers via certified mail on 8/22/08 informing them of the problem. Until the exchange of affected printed circuit boards has been implemented, the firm is recommending that users closely monitor all device displays and error messages and if applicable act in accordance with the advisory statements and remedial information provided in the Instructions for Use as well as making provisions for alternative patient care measures.

Root Cause

Nonconforming Material/Component

Distribution Pattern

Nationwide Distribution.