Terminated

chromID Salmonella Agar, Reference No. 43621. A selective isolation and differentiation medium for the detection of Salmonella in human specimens.

FDA Recall Z-0087-2014 · Posted 2013-10-28

Recalling FirmbioMerieux, Inc.
Recall StatusTerminated
Date Initiated2013-09-06
Date Terminated2014-06-24
Product Quantity90

Reason for Recall

Complaints for Salmonella strains growing on this medium producing an uncharacteristic white colored colony instead of the expected pale pink to mauve colored colonies. This could potentially lead to Salmonella colonies not being detected.

Firm Action

Recall letters were issued to the US customers by Biomeriux, Durham, NC on 9/6/2013.

Root Cause

Nonconforming Material/Component

Distribution Pattern

Nationwide Distribution.