Terminated

da Vinci Si Surgical System IS3000 with the da Vinci Surgeon Console which contains a high Resolution Stereoscopic Viewer. Intended to assist in the accurate control of Intuitive Surgical Endoscopic instruments.

FDA Recall Z-0088-2014 · Posted 2013-10-28

Recall StatusTerminated
Product CodeNAY
Date Initiated2011-10-17
Date Terminated2013-10-28
Product Quantity1,090 consoles

Reason for Recall

Improper restraints during transportation of the da Vinci system could cause the HRSV Monitor to become loose from its mount.

Firm Action

Advisory letters were sent to all affected customers on 10/17/11 informing them to contact Customer Service prior to moving the system outside of the hospital.

Root Cause

Use error

Distribution Pattern

Worldwide Distribution-USA (nationwide) including Puerto Rico and the countries of Australia, Austria, Belgium, Canada, Chile, China, Denmark, Egypt, Finland, France, Greece, Germany, India, Indonesia, Israel, Ireland, Italy, Kuwait, Mexico, Netherlands, Norway, Philippines, Poland, Portugal, Qatar, Romania, Saudi Arabia, Singapore, Slovakia, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, and United Kingdom.