Terminated

SDMS (SenTec Digital Monitoring System)

FDA Recall Z-0088-2017 · Posted 2016-10-17

Recalling FirmSenTec AG
Recall StatusTerminated
Product CodeLKD
Date Initiated2016-09-15
Date Terminated2017-06-23
Product Quantity75 (54 US; 21 =OUS)

Reason for Recall

Sentec AG found that with one batch of an internal SenTec Monitor component (Docking Station Module) there is a risk that a fastening screw could become loose under certain circumstances. A loose screw within the monitor can create uncontrolled electrical connections, including short circuits, with the risk of smolder or fire and electrical shock (patient and/or operator).

Firm Action

Distributors were sent via e-mail a SenTec 'Mandatory initiated field correction action" letter dated September 15, 2016 . The letter provided a "Description of the issue, Affected products, Identify and localize affected products, Removal from use, Replacement of affected products at end customer & distributor / sub-distributor level". It also provided "action information for sub-distributors and affiliates, and end customers". For questions contact SenTec Inc. by e-mail ([email protected]) or phone (636-343-0000).

Root Cause

Component change control

Distribution Pattern

Worldwide Distribution - US Distribution to the states of : LA, MI, MO, OH, TN. and to the countries of : United Kingdom, The Netherlands and Australia.

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.

Official Source

View FDA Recall →