Premicath 28G PICC Catheter
FDA Recall Z-0090-06 · Posted 2005-10-28
Advertisement
Recalling FirmVygon Corporation
Recall StatusTerminated
Product CodeLJS
Date Initiated2005-06-27
Date Terminated2005-10-31
Product Quantity243 units
Reason for Recall
Guidewire protuding from tip of the catheter
Firm Action
The recalling firm issued a recall letter via fax to the distributors on 6/27/05. The letter informed the distributors of the problem and to notify any hospital accounts of the problem and the need to trim the catheter prior to implantation.
Root Cause
Other
Distribution Pattern
The product was shipped to distributors in CA, CO, FL, IL, MD, NJ, NY, OH, PA, TX and WA.