ConMed DetachaTip Laparoscopic Instrument: Fenestrated (Duckbill) Multiple Use Grasper, 5mm x 43cm, REF/Product Code 1-4307. --- The product is distributed sterile in heat-sealed tray and labeled with an expiration date that pertains only to the sterility.

FDA Recall Z-0090-2007 · Posted 2006-10-27

Recalling FirmConmed Corporation
Recall StatusTerminated
Product CodeGEI
Date Initiated2006-10-12
Date Terminated2012-08-30
Product Quantity508 units

Reason for Recall

The grasper jaws broke during laparoscopic procedures at the junction of the jaw and the tube.

Firm Action

On 10/12/06, ConMed sent "URGENT DEVICE RECALL" packages (letter dated 10/11/06) to the domestic accounts via UPS Ground, and via UPS Express for the International accounts. Devices are to be removed from use and returned.

Root Cause

Nonconforming Material/Component

Distribution Pattern

Worldwide