Reprocessed Agilis Steerable Introducer: indicated for introducing various cardiovascular catheters into the heart. (a) small curl (16.8mm curve type/71 cm length), Product Code STJ408309; (b) medium curl (22.4 mm curve type/71 cm length), Product Code STJ408310; (c) large curl (50 mm curve type/71 cm length), Product Code STJ408324
FDA Recall Z-0090-2019 · Posted 2018-10-09
Advertisement
Recalling FirmSterilmed, Inc.
Recall StatusTerminated
Product CodePNH
Date Initiated2018-09-17
Date Terminated2020-08-04
Product Quantity218 units
Reason for Recall
Reprocessed Agilis Steerable Introducers may not meet endotoxin requirements per product specification for biological residues determined by LAL testing.
Firm Action
Sterilmed notified customers on about 09/17/2018, via URGENT MEDICAL DEVICE RECALL letter sent via UPS next day. Instructions included examine inventory for affected devices, remove and return affected devices to Stericycle (3rd party processor), provide recall information to appropriate personnel at the facility, and complete and return the Business Reply Form to Stericycle. For questions, contact Sterilmed's Customer Care Team at (888) 541-0078, Monday through Friday from 7:00 AM to 5:00 PM CST.
Root Cause
Material/Component Contamination
Distribution Pattern
U.S. Nationwide distribution in the states of CA, CO, MN, and WA.
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.