AQUIOS CL Flow Cytometer, Catalog No. B30166, Rx. The firm name on the label is Beckman Coulter Ireland Inc., Ireland. The AQUIOS CL Flow Cytometer is intended for use as an in vitro diagnostic device for immunophenotyping. It is used in conjunction with the following reagents and software package: AQUIOS Tetra-1 Panel monoclonal antibody reagents, AQUIOS Tetra-2+ Panel monoclonal antibody reagents, and AQUIOS CL Flow Cytometry Software.

FDA Recall Z-0091-2018 · Posted None

Recalling FirmBeckman Coulter Inc.
Recall StatusTerminated
Product CodeOYE
Date Initiated2017-10-02
Date Terminated2019-05-08
Product Quantity268 units

Reason for Recall

The system may duplicate sample requests leading to sample misidentification when connected to a Laboratory Information System (LIS)

Firm Action

Beckman Coulter sent an Urgent Medical Device Recall letter dated September 28, 2017, and via phone on October 4, 2017. On 11/30/2017, the firm issued an updated letter dated 11/28/2017. The letter informs the customer this letter replaces the instructions in the original notification dated 9/28/2017.

Root Cause

Software design

Distribution Pattern

Worldwide Distribution - US Distribution to the states of : AZ, CA, CO, FL, GA, IL, IN, KS, NC, NY, OR, PA, SC, TX, and WA, including Puerto Rico and government distribution, and to the countries of : Canada, Angola, Australia, Belgium, Botswana, Brazil, Chile, Colombia, Czech Republic, Denmark, Finland, France, Germany, Ghana, Hong Kong, India, Israel, Italy, Kenya, Lesotho, Malawi, Malaysia, Mexico, Namibia, Netherlands, New Zealand, Norway, Pakistan, Panama, Poland, Portugal, Romania, Slovakia, South Africa, Spain, Sweden, Switzerland, Turkey, Uganda, Ukraine, United Kingdom, and Zambia.