DePuy Synthes Radiolucent Retractor, part number 387.580, lot number 3620715
FDA Recall Z-0092-2015 · Posted 2014-10-20
Advertisement
Recalling FirmSynthes, Inc.
Recall StatusTerminated
Product CodeGAD
Date Initiated2014-10-02
Date Terminated2015-12-18
Product Quantity11
Reason for Recall
Particle residue on the instrument from adhesive tape which was used to bind the device during transport. There may be a risk of tissue reaction if a patient is exposed to latex particles from the adhesive tape, or a systemic reaction such as anaphylaxis in patients with latex allergy.
Firm Action
DePuy Synthes sent an Urgent Notice Medical Device Recall letter dated October 2, 2014, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer.
Please take the following actions:
If you DO have any of the identified devices, please take the following steps:
Call DePuy Synthes at 1-800-479-6329 to obtain a Return Authorization (RA) Number.
Complete the Verification Section (page 3 of this letter) by checking the appropriate box indicating affected product has been located. Also, please indicate the number of devices found and note the Return Authorization Number. Please include your name, title, address, telephone number and signature in the spaces provided.
Send a copy of the completed Verification Section to DePuy Synthes by:
Fax: (610) 430-7083 or " Scan/email: [email protected]
Note: If the Verification Section is answered on behalf of more than one facility and/or individual, please clearly indicate the name and address of the facility and/or individual on page 3 of the notification.
If you DO NOT have the identified product, please take the following steps:
Complete the attached Verification Section (page 3 of this letter) by checking the appropriate box indicating that no affected product has been located. Please include your name, title, address, telephone number and signature in the spaces provided. This return documentation acknowledges your receipt of medical device removal information.
Return the documents to DePuy Synthes by Fax: (610) 430-7083 or
Scan/email: [email protected]
Note: If the Verification Section is answered on behalf of more than one facility and/or individual, please clearly indicate the name and address of the facility and/or individual on page 3 of the notification.
If you have any questions, please call 610-719-5450, or contact your DePuy Synthes sales consultant. Thank you for your attention to this issue.
Root Cause
Material/Component Contamination
Distribution Pattern
US Distribution including the states of PA, NC, TN, NY, WI, WA and MI.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.