Disclaimer: This site aggregates publicly available data from official government sources (FDA, ClinicalTrials.gov, PubMed, SEC EDGAR) for general reference only. It does NOT constitute medical advice, diagnosis, treatment recommendations, or investment advice.

Terminated

AG Therapist 3rd party V&R equipped with a Coherence Therapist R2.0 or R2.1, and PRIMEVIEW 3iR2.0 or R2.1, part number 7345411 Medical charged-particle radiation therapy system.

FDA Recall Z-0093-2009 · Posted 2008-10-16

Recall StatusTerminated
Product CodeIYE
Date Initiated2007-11-06
Date Terminated2010-12-18
Product Quantity1 unit

Reason for Recall

Flat panel positioning calibration could be off by as much as 4 mm without the machine discovering detail.

Firm Action

Service calibration inspection/correction Update instructions TH038/07/S was issued on 11/06/2007 to all customers. The instructions identified the affected products, stated the reason for the update, and listed the steps to update the software. Customer notification was also issued. Software patch planned for release in June 2008.

Root Cause

Software design

Distribution Pattern

Worldwide Distribution

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.

Official Source

View FDA Recall →

Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.

↑