Terminated

McKesson Cardiology Hemo. McKesson Cardiology" Hemo is intended for complete physiological/hemodynamic monitoring, clinical data acquisition, medical image and data processing, and analytical assessment.

FDA Recall Z-0093-2015 · Posted 2014-10-20

Recall StatusTerminated
Product CodeMWI
Date Initiated2014-06-25
Date Terminated2017-10-26
Product Quantity625 devices

Reason for Recall

McKesson Cardiology Hemo front-end parameter boxes must undergo annual maintenance testing to establish the proper functioning of the parameter boxes, and to identify in a timely manner whether a parameter box is out of calibration. McKesson has identified that its communication to customers regarding the required maintenance tests was insufficient. While each unit has an attached sticker i

Firm Action

Consignees were notified via Field Safety Notice on 06/25/2014.

Root Cause

Labeling design

Distribution Pattern

Worldwide Distribution-USA (nationwide) and the countries of Australia, Botswana, Canada, Germany, New Zealand, Netherlands, Pakistan, Portugal, South Africa, Turkey and The United Kingdom.