Disclaimer: This site aggregates publicly available data from official government sources (FDA, ClinicalTrials.gov, PubMed, SEC EDGAR) for general reference only. It does NOT constitute medical advice, diagnosis, treatment recommendations, or investment advice.

Open, Classified

MED-810A Zemits NDPrime Laser products

FDA Recall Z-0095-2024 · Posted 2023-10-13

Recalling FirmAdvance-Esthetic LLC
Recall StatusOpen, Classified
Product CodeGEX
Date Initiated2023-10-09
Product Quantity19

Reason for Recall

Advance-Esthetic LLC failed to comply with the applicable standards regarding Electronic Product Radiation Control (EPRC) regulations associated with the laser product performance standards and certification. This Recall provides correction by the firm regarding Zemits NdPrime Yag Tattoo Removal Laser. The product requires update or is missing the following in part: remote interlock connector, safety interlocks, manual reset mechanism, beam attenuator, tests for determination of compliance, certification, class IV designation and warning, labels, and labeling.

Firm Action

Notification letters will be sent to affected customers identifying the reason for notification and action to be taken. The manufacturer will, without charge, remedy the defect or bring the product into compliance with each applicable Federal standard in accordance with a plan to be approved by the Secretary of Health and Human Services. For any queries or require further assistance, our customer support team is available at 888-999-3996 or [email protected].

Root Cause

Radiation Control for Health and Safety Act

Distribution Pattern

US Nationwide Distribution

Official Source

View FDA Recall →

Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.

↑