ENDOVASCULAR PACK; Model/Item Number (Kit): VAPX014-04;
FDA Recall Z-0095-2027 · Posted 2026-10-07
Advertisement
Recalling FirmAVID Medical, Inc.
Recall StatusOpen, Classified
Product CodeOEZ
Date Initiated2026-08-10
Product Quantity33
Reason for Recall
Product may be potentially contaminated with a lubricant used during the manufacturing process.
Firm Action
On August 10, 2026, URGENT: MEDICAL DEVICE RECALL letters were sent to customers.
Actions to be taken:
1. Identify, segregate, and quarantine all Avid kits identified in Attachment 1.
2. Add recall labels to affected Avid kits to ensure all users are aware of the actions to discard the impacted component. A labeling template is enclosed for printing labels which can be affixed on all affected kits and cases in inventory. Please contact your Sales Representative if pre-printed labels are needed. This label should be applied in a prominently visible location to end-users. The location chosen should not cover any other critical product information found on existing product labeling.
- Customers (end-users): If you received product from a distributor, please contact the distributor or your sales representative with any questions you may have.
-Distributors: If any of the affected kits identified above were further distributed, please ensure all end-users are appropriately notified and maintain records of effectiveness.
Root Cause
Process change control
Distribution Pattern
US Nationwide distribution in the states of AZ, CA, FL, MD, NE, TX.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.