smith&nephew TRIGEN INTERTAN 10S, REF 71675384, 10 MM X 18 CM NAIL, 130o (degrees), SET SCREW PREASSEMBLED IN NAIL
FDA Recall Z-0097-2018 · Posted None
Advertisement
Recalling FirmSmith & Nephew, Inc.
Recall StatusOpen, Classified
Product CodeJDS
Date Initiated2017-10-24
Product Quantity17 units
Reason for Recall
A single lot of INTERTAN 10S 10 MM X 18 CM 130 D was manufactured with the screw inserted upside down.
Firm Action
Smith & Nephew sent an Urgent Medical Device Recall Notice dated October 24, 2017, to all affected customers. The firm requested return of the product.
Please follow the instructions on the attached Response Form. For further questions, please call (978) 749-1440.
Root Cause
Employee error
Distribution Pattern
US Distribution - TN, IL, CT, MO, CA., and Internationally to Italy and Germany