Clea Stand Ceiling used with the Integris X-Ray System H5000, Allura 12/15, Allura 9C

FDA Recall Z-0099-04 · Posted 2004-07-20

Recall StatusTerminated
Product CodeEHD
Date Initiated2003-10-10
Date Terminated2005-01-13
Product Quantity536 units

Reason for Recall

The cable carrier and cables may fall after repeat collisions with OR lamps, radiation shields or other supported accessories.

Firm Action

The firm sent a letter dated, October 10, 2003 indicating the firm''s service representatives will upgrade the cable carrier by mounting a safety catch inside the cable carrier.

Root Cause

Other

Distribution Pattern

The firm distributes to hospitals and medical centers nationwide.