Portex Perfit Percutaneous Tracheostomy Tube Replacement Kit, Sterile Ref: 536090

FDA Recall Z-0100-2008 · Posted 2007-11-07

Recall StatusTerminated
Product CodeJOH
Date Initiated2007-09-18
Date Terminated2012-05-08
Product Quantity40 units

Reason for Recall

Mislabeled. The product is labeled as a size 9.0 mm tracheostomy tube, but the package contains a 7.0 mm tracheostomy tube.

Firm Action

Smiths Medical contacted direct customers and US distributors by letter sent by UPS on Septmeber 18, 2007. US Distributors instructed to return their inventory and provide a customer list to Smiths Medical.

Root Cause

Process control

Distribution Pattern

Nationwide.