Terminated

Vari-Lase Endovenous Laser Procedural Kit, Model 7000

FDA Recall Z-0101-04 · Posted 2003-11-06

Recall StatusTerminated
Product CodeGEX
Date Initiated2003-10-09
Date Terminated2005-11-21
Product Quantity390 kits

Reason for Recall

An internal gasket of the Fiber Lock component of the Vari-Lase Kit contains natural rubber latex, but the product labeling does not recognize the presence of latex and include the required cautionary statement.

Firm Action

Letters dated October 9, 2003 were sent to the consignees. The letters instruct the consignees to place the enclosed warning stickers, about latex in the product, on the boxes and pouches of any remaining product in inventory at the consignees. After completing the relabeling, consignees are to fill out the attached form and fax it to the firm to document the field correction.

Root Cause

Other

Distribution Pattern

The product was distributed nationwide in the United States and to Romania.