Terumo Advanced Perfusion System 1 Roller Pump; 4 inch diameter; Catalog number 801040.

FDA Recall Z-0102-04 · Posted 2003-11-07

Recall StatusTerminated
Product CodeDTQ
Date Initiated2003-10-16
Date Terminated2005-07-07
Product Quantity285

Reason for Recall

The roller pump tube clamp mechanism may fail to release , making it difficult to remove or insert the tubing, and thereby delaying perfusion.

Firm Action

Letters entitled Urgent Safety Alerts were sent to each customer on 10/16/03. Customers were made away of the problem and advised to develop a protocol for action, in case the problem occurred in their facility.

Root Cause

Other

Distribution Pattern

United States, Belgium, Canada, Dubai, Egypt, Germany, Hong Kong, India, Indonesia, Japan, Malaysia, Singapore, South Korea.