Disclaimer: This site aggregates publicly available data from official government sources (FDA, ClinicalTrials.gov, PubMed, SEC EDGAR) for general reference only. It does NOT constitute medical advice, diagnosis, treatment recommendations, or investment advice.

Terminated

Biomet Orthopedics Apical Hole Plug, 3/8 - 24 threaded, TI-6AL-4V alloy, sterile; REF 123741, Biomet Orthopedics, Inc., Warsaw, IN 46581

FDA Recall Z-0102-2008 · Posted 2007-10-27

Recalling FirmBiomet, Inc.
Recall StatusTerminated
Product CodeLPH
Date Initiated2007-08-09
Date Terminated2009-06-26
Product Quantity5

Reason for Recall

Incorrect or lack of threading: The plug component was manufactured incorrectly and does not fit with the screw hole as intended.

Firm Action

The distributor was notified via phone on or about 8/9/07 and by letter dated 8/31/07 and requested to return the product. Returns included subaccounts

Root Cause

Process control

Distribution Pattern

Nationwide

Official Source

View FDA Recall →