SoftPath GUI version 4.2.3.4 2008-06-30 (Manuf Date). SoftPath GUI version 4.2.4.1 2008-09-30 (Manuf Date). SoftPath GUI version 4.3.0.3, 2008-07-23 (Manuf Date). SoftPath GUI version 4.3.0.4 2008-09-16 (Manuf Date). SCC Soft Computer, 5400 Tech Data Drive, Clearwater, FL 33760.
FDA Recall Z-0102-2011 · Posted 2010-10-20
Advertisement
Recalling FirmSCC Soft Computer
Recall StatusTerminated
Product CodeLNX
Date Initiated2009-01-26
Date Terminated2010-11-03
Product Quantity52
Reason for Recall
For SoftPath GUI versions 4.2.3.4, 4.2.4.1, 403.0.3, and 4.3.0.4., the Previous Patient History and some QA Management reports may have missing or incorrect diagnosis text. However, the patient final report, HIS resulting and Query display are all correct. All affected clients have been provided corrected software..
Firm Action
SCC Soft Computer issued a Risk-to-Health notification with Client Notification Verbiage to affected clients using SCC Soft Computer's proprietary Task Management System on January 26, 2009. SCC issued a software upgrade with corrective patch electronically to all clients. Client was asked to acknowledge receipt of the task and grant permission to load the required software.
SCC will contact each client to coordinate the running of the utility and the loading and validation of the correct patch.
Root Cause
Software design
Distribution Pattern
Worldwide Distribution: Throughout USA and Canada.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.