Pointe Scientific Liquid Glucose (HEXO) Reagent Set, Pointe Scientific, Inc, Canton, MI; Catalog #'s G7517-120,G7517-120, G7517-500, and G7517-1L; 8-G7517-120, 8-G7517-500, 3-G7517-L, 7-G7517-500 and 7-G7517-1000. (Clinical chemistry) Pointe Scientific Liquid Glucose (HEXO) is used for the quantitative determination of glucose in serum.
FDA Recall Z-0104-2010 · Posted 2009-11-04
Advertisement
Recalling FirmPointe Scientific, Inc.
Recall StatusTerminated
Product CodeCFR
Date Initiated2009-10-19
Date Terminated2010-09-27
Product Quantity400 kits plus 59 liters and 15 vials.
Reason for Recall
There is a loss of linearity. The product fails to maintain linearity specification of 500 mg/dL.
Firm Action
The firm issued a press release, and an amended press release, on 10/30/09. Consignees were notified by an Urgent: Medical Device Recall letter dated 10/19/09, which identified the problem; and, instructed testing laboratories to inform patients' attending physicians, and determine, with their input, whether confirmation of the previous test results will be required. Customers are to examine their inventory for any of the lot numbers listed on the letter; and, for any product that has been further distributed, those customers need to be identified and notified of this product recall by including a copy of the recall letter. Customers are to complete and return the Recall Return Response form and upon receipt, Pointe Scientific will replace any disposed of reagent. The recall was expanded to include two additional lots by letter dated 11/2/09.
Root Cause
Process design
Distribution Pattern
Worldwide Distribution -- USA, Israel, Mexico, South Africa, and West Indies.