HARVEST(R) TERUMOBCT, BMAC2-120-01Bone Marrow Procedure Pack, REF 51423, Rx Only, STERILE EO

FDA Recall Z-0104-2018 · Posted None

Recalling FirmTerumo BCT, Inc.
Recall StatusTerminated
Product CodeJQC
Date Initiated2017-10-20
Date Terminated2020-08-21
Product Quantity281 units

Reason for Recall

The patient labels included in the BMAC system procedure packs, catalog numbers 51416 and 51423, are printed with an incorrect lot number, catalog number, and expiry date; the lot number on the procedure pack box label is correct. The expiration noted was not beyond initiation date of the recall. Since the product is noted as sterile, the clearance for K103340, K052925 and K991430 raises question if product was cleared with a sterility claim. Further post-market follow-up will be required to assess the safety risk of the product.

Firm Action

The firm initiated their recall by letter on 10/20/2017. The letter stated the following: "ACTIONS REQUIRED FOR HEALTHCARE PROVIDERS AND DISTRIBUTORS 1. For product shelved in the original procedure pack (white box), examine the label for catalog/lot numbers as shown in Figure C above. 2. If affected product is in your inventory, contact the Terumo BCT Customer Support Center at +1.877.3.FYI.BCT (U.S. Toll-Free +1.877.339.4228) or +1.303.231.HELP (+1.303.231.4357), Terumo BCT Europe N.V. at +32.2.715.0590, or your local Terumo BCT representative, and Terumo BCT will issue a Return Goods Authorization for product return. 3. Continue to use your unaffected product in accordance with the instructions for use. 4. Distribute this notification to all Harvest Bone Marrow Aspirate Concentrate (BMAC) System users within your organization. 5. IMPORTANT: Complete the attached acknowledgement and return it by fax or email to Terumo BCT by 27 October 2017. Your return of the acknowledgement is critical so we can confirm that you have received the recall notice.

Root Cause

Process control

Distribution Pattern

CA, Co, MO, NY, PA, TX, VA, WI, Turkey, Austria, Spain