Chesapeake Anterior Lumbar (AL) Removal Tool; Cat. no. 2008-90068
FDA Recall Z-0104-2022 · Posted None
Advertisement
Recalling FirmK2M, Inc
Recall StatusTerminated
Product CodeOVE
Date Initiated2021-09-08
Date Terminated2024-05-17
Product Quantity23
Reason for Recall
Units may arrive with missing pins and/or in a state in disassembly. Pins dislodging from the instrument during a procedure may enter the surgical cavity, requiring intraoperative removal or revision surgery.
Firm Action
On September 8, 2021, the firm distributed Urgent Medical Device Recall letters to affected customers.
Actions Needed:
1. Immediately check your internal inventory to locate the product listed in the letter and remove them from their point of use.
2. Use the included Business Reply Form to reconcile any affected product. Complete the Business Reply Form even if there is no affected product identified.
3. Return the enclosed Business Reply Form by email to [email protected] to confirm receipt of this notification/document product segregation. Your account s Stryker Spine Sales Representative will arrange for retrieval and removal of any instruments associated with this recall.
4. In the interim, until units can be removed from your facility by Stryker, there are no additional actions that users can and should take once the product has been segregated and removed from point of use.
- If you desire additional training associated with these instructions, please contact our Regulatory Compliance team by calling 201.749.8090 or by email at [email protected], or contact your local Sales representative.
5. Maintain awareness of this communication internally until all required actions have been completed within your facility.
6. Inform Stryker if any of the subject devices have been distributed to other organizations. If so, provide contact details so Stryker can inform the recipients appropriately.
7. Under 21 CFR 803, manufacturers are required to report any serious injuries where a product has contributed or may have contributed to the event. Please keep Stryker informed of any adverse events associated with this product by using Stryker s online reporting site: https://www.stryker.com/productexperience. Additionally, adverse reactions or quality problems experienced with the use of this product may also be reported to the FDA's MedWatch Adverse Event Reporting program: https://www.accessdata.fda.gov/scripts/medwatch/index
Root Cause
Process control
Distribution Pattern
Domestic distribution to CO, IL, LA, MA, MS, OK, PA, TN, and TX. Foreign distribution to Columbia, UK, Netherlands, and Canada.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.