Terminated

Siemens Models: BIOGRAPH, BIOGRAPH LSO and CTI Models: REVEAL HD, REVEAL RT

FDA Recall Z-0106-04 · Posted 2003-11-13

Recalling FirmCti Pet Systems Inc
Recall StatusTerminated
Product CodeKPS
Date Initiated2003-10-02
Date Terminated2006-02-07
Product Quantity73

Reason for Recall

CPS has found that in some instances, when the CT E-Stop is activated, a malfunction may occur where the PHS controller may not shut down and stop bed motion.

Firm Action

Written notification was sent out beginning 10/02/2003 to distributors and end users to advise them of the problem.

Root Cause

Other

Distribution Pattern

Nationwide International: Japan, Germany, Brazil, Spain, China, Russia, Austrailia, Hong Kong, Singapore, Poland France, Korea, Greece, UK and Austria