Terminated

Box Label: 3 mm Cueva Electrode Kit, Sterile EO, Rx Only, REF. CNE1-3KDINX . Pouch labels: Each kit contains 3 pouches. 3 mm Cueva Electrode CNE-3X, Cueva Electrode Applicator Wand CNE-100X, Cueva Electrode Leadwire CNE-LW-2DINX. Box Label: 2 mm Cueva Electrode Kit, Sterile EO, Rx Only, REF. CNE1-2KDINX . Pouch labels: Each kit contains 3 pouches. 3 mm Cueva Electrode CNE-2X, Cueva Electrode Applicator Wand CNE-100X, Cueva Electrode Leadwire CNE-LW-2DINX.

FDA Recall Z-0106-2015 · Posted 2014-10-21

Recall StatusTerminated
Product CodeGZL
Date Initiated2014-10-08
Date Terminated2015-04-08
Product Quantity13 kits (26 electrodes, 26 applicator wands, 26 leadwires).

Reason for Recall

The recall has been initiated due to concerns that the applicator wand may malfunction potentially causing the inability to monitor the VII cranial nerve. .

Firm Action

Consignees were sent on 10/8/2014 an AD-TECH "Medical Device Recall" letter dated October 6, 2014. The letter described the Reason for the Voluntary Recall, Risk to Health, Action to be taken by the Customer, Product and Distribution Information and Other Information. Requested consignees to return the "Acknowledgement and Receipt Form" to their Ad-Tech Clinical Specialists Fax 262-634-5668, Telephone 262-634-1555, [email protected]. For additional information they can use the same number.

Root Cause

Process control

Distribution Pattern

Worldwide Distribution: US (nationwide) in the states of: CO, FL, LA, NJ, OH, WA, and WI; and internationally to: Finland.