Ion Flexison Biopsy Needle, 19G, Extension Range : 0-3 cm, Ref: 490104-09 and 49004-10, 5-pack, Sterile EO, Rx only, UDI:00886874115978

FDA Recall Z-0106-2022 · Posted None

Recall StatusTerminated
Product CodeEOQ
Date Initiated2021-08-06
Date Terminated2024-05-17
Product Quantity1625 units

Reason for Recall

Potential for the sheath tip on the biopsy needle to separate from the sheath shaft.

Firm Action

On August 6, 2021, Intuitive Surgical issued an "Urgent Medical Device Recall" notice to affected consignees via Fed Ex. In addition to informing consignees about the recall, Intuitive ask consignees to take the following actions: 1. Inform necessary hospital personnel about this issue. 2. Inform affected personnel when the requested actions have been completed. 3. Complete the attached Acknowledgement Form immediately and return it via fax or email to Intuitive as instructed on the form. 4. Please retain a copy of this letter and the acknowledgement form for your files. 5. All returned affected product will be assessed and credit will be provided for the quantity returned. Intuitive will provide a follow up notification once unaffected product becomes available. 6. If you need further information or support concerning this Medical Device Notification, please contact your Endoluminal Sales Representative or contact Intuitive Customer Service at the numbers listed below: " North America: (800) 876-1310, Option 3 (4 AM to 5 PM PST) or mail: [email protected].

Root Cause

Under Investigation by firm

Distribution Pattern

US: AZ, AR, CA, CO, DC, FL, GA, IL, IN, IO, MD, MA, MI, MN, MO, NV, NJ, NY, NC, OH, OK, PA, TN, TX, UT, VA, WA OUS: China