SPECTRUM IQ INFUSION SYSTEM WITH DOSE IQ SAFETY SOFTWARE, 3570009

FDA Recall Z-0107-2019 · Posted None

Recall StatusTerminated
Product CodePHC
Date Initiated2018-09-18
Date Terminated2020-08-19
Product Quantity354 units

Reason for Recall

If an operator attempts to manually program a Basic Infusion for a drug that is not included in the Drug Library using the Basic-Dose Mode, the operator will receive a false hard limit exceeded error, preventing him or her from programming the infusion.

Firm Action

The firm initiated their field correction by email on 09/14/2018. The notice alerted the consignee of the problem and stated that Baxter would be issuing a software update to correct the issue.

Root Cause

Software design

Distribution Pattern

NJ