CADD-Solis VIP ambulatory infusion pump, Model Numbers: a. 21-2120-0105-02L b. 21-2120-0105-03L c. 21-2120-0105-06L d. 21-2120-0105-07L e. 21-2120-0105-08L f. 21-2120-0105-12L g. 21-2120-0105-13L h. 21-2120-0105-14L i. 21-2120-0105-15L j. 21-2120-0105-17L k. 21-2120-0105-50L l. 21-2127-0105-02L m. 21-2127-0105-06L n. 21-2127-0105-50L

FDA Recall Z-0107-2022 · Posted 2021-10-19

Recall StatusTerminated
Product CodeFRN
Date Initiated2020-03-24
Date Terminated2023-12-22
Product Quantity642 units

Reason for Recall

Specific Pumps may exhibit intermittent performance in the AILD (Air in Line Detector) function. If an AILD does not sufficiently discern fluid from air in line, an air in line event may not be recognized by the pump and may not alarm to notify the clinician.

Firm Action

The firm notified its consignees by email beginning 03/24/2020. The notice explained the issue, risk to health, and requested a response. Distributors/pharmacy suppliers were directed to notify their customers. The firm is seeking return of the units for correction.

Root Cause

Component change control

Distribution Pattern

Distribution to European countries: BE, BH, CH, DE, DK, ES, FI, FR, GB, IT, HL, NO, PT, SE