Flow Re-Direction Endoluminal Devices: FRED 27, REF: FRED5019, FRED5029, FRED4528, FRED4026, FRED5526, FRED4508, FRED5514, FRED3536, FRED3511, FRED5014, FRED3516, MV-F451827; FRED X 27, REF: XFRED4017-MVE, XFRED4528-MVE, MV-F501427X, XFRED4518-MVE, XFRED5526-MVE, XFRED4539-MVE, FREDX5522-PMA, FREDX3522-PMA, FREDX4525-PMA, FREDX5015-PMA, FREDX5532-PMA, FREDX4013-PMA, FREDX3513-PMA, FREDX4515-PMA, FREDX3517-PMA, MV-F352427X

FDA Recall Z-0108-2026 · Posted 2025-10-10

Recalling FirmMICROVENTION INC.
Recall StatusOpen, Classified
Product CodeOUT
Date Initiated2025-09-05
Product Quantity467

Reason for Recall

Flow diverter may have a manufacturing issue related to the tantalum length and/or tantalum attachment pattern, which may lead the device to be incompletely open or not properly apposed to the vessel wall, which may necessitate device recapture and withdrawal of the delivery system and microcatheter together. Insufficient apposition could lead to long-term health consequences.

Firm Action

On 9/5/2025, recall notices were mailed to customers who were asked to do the following: 1) Stop using and quarantine affected devices. 2) Provide the notice to medical facilities or users to whom you have distributed recalled devices. 3) Return effected devices. 4) Complete and return the customer acknowledgement form via email to [email protected] Direct questions to the firm at [email protected] On 10/29/2025, additional recall notices were distributed identifying additional distributed devices.

Root Cause

Device Design

Distribution Pattern

Worldwide - US Nationwide distribution in the states of SD, CA, SC, TX, FL, CO, NE, NY and the countries of DE, ES, PL, GB, IT, IN, AU, KZ, JP, AT, SI, RS, GR, TR, AR, CH, SG, KR, SK, MX, DK, IR, SA, BY, PA, VN, ZA, BR, TW, CN, BE, FR, VE, MY.