Boston Scientific-Renegade Hi-Flo Microcatheter Kit, UPN Material Number: MOO1183030,Catalog Number: 18-303

FDA Recall Z-0109-2008 · Posted 2007-10-31

Recall StatusTerminated
Product CodeKRA
Date Initiated2007-09-10
Date Terminated2007-12-20
Product Quantity153

Reason for Recall

Incorrect Guidewire: Product may incorrectly contain a 135 cm length Transend-18 guidewire instead of the kit labeled 165 cm length Transend-18 guidewire.

Firm Action

Boston Scientific issued a Urgent Medical Device Recall Letter dated September 10, 2007 to Domestic accounts via Federal Express. The notification instructed the account to immediately discontinue the use of the product under this recall, check their inventory, remove any affected product , and return it to Boston Scientific. The notification also requested the completion of a Reply Verification Tracking Form

Root Cause

Packaging process control

Distribution Pattern

Nationwide