Boston Scientific-Renegade Hi-Flo Microcatheter Kit, UPN Material Number: MOO1183030,Catalog Number: 18-303
FDA Recall Z-0109-2008 · Posted 2007-10-31
Advertisement
Recalling FirmBoston Scientific Corporation
Recall StatusTerminated
Product CodeKRA
Date Initiated2007-09-10
Date Terminated2007-12-20
Product Quantity153
Reason for Recall
Incorrect Guidewire: Product may incorrectly contain a 135 cm length Transend-18 guidewire instead of the kit labeled 165 cm length Transend-18 guidewire.
Firm Action
Boston Scientific issued a Urgent Medical Device Recall Letter dated September 10, 2007 to Domestic accounts via Federal Express. The notification instructed the account to immediately discontinue the use
of the product under this recall, check their inventory, remove any affected product , and return it to Boston Scientific. The notification also requested the completion of a Reply Verification Tracking Form
Root Cause
Packaging process control
Distribution Pattern
Nationwide
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.