Fujinon/Fujifilm Ultrasonic Endoscope Intended to provide ultrasonic images of submucosal and peripheral organs of the upper gastrointestinal tract for observation, diagnosis and endoscopic treatment.

FDA Recall Z-0112-2014 · Posted 2013-10-30

Recalling FirmFujinon Inc
Recall StatusTerminated
Product CodeFDS
Date Initiated2013-07-31
Date Terminated2014-03-04
Product Quantity12 Units

Reason for Recall

A limited number of endoscopes were shipped to FMSU-ESD with a manufacturing defect which included missing set screws used to more securely attach the Forceps Inlet (biopsy inlet port) to the endoscope.

Firm Action

The firm, FUJIFILM Medical Systems, USA, Inc. (FMSU-ESD) sent an "URGENT-MEDICAL DEVICE CORRECTION (UMDC)" and Customer Acknowledgement forms dated July 31, 2013 to their customers. The letter described the issue with the affected product; instructed customers on how to identify affected units; instructed customers to immediately cease use of the affected units; return affected units to FMSU-ESD for repairs and quality control inspection; and complete and return the attached "Customers Acknowledgement" form to FMSU-ESD, Attn: Regulatory Compliance, via fax to (973) 686-2616 or email: [email protected]. If you have any questions regarding this notice, please do not hesitate to contact Director, Quality and Regulatory Compliance, at (973) 686-2636 or by email: [email protected].

Root Cause

Unknown/Undetermined by firm

Distribution Pattern

Worldwide distribution: US (nationwide) including states of: NJ, PA and VA; and country of Canada.