Terminated

Monaco Radiation Treatment Planning System. Used to make treatment plans for patients with prescriptions for external beam radiation therapy.

FDA Recall Z-0112-2016 · Posted 2015-10-07

Recalling FirmElekta, Inc.
Recall StatusTerminated
Product CodeMUJ
Date Initiated2015-09-15
Date Terminated2017-04-04
Product Quantity154 units

Reason for Recall

Dose and MU are incorrect when CT images are viewed from the head, and, when using multiple prescriptions with forced densities.

Firm Action

The firm issued an URGENT IMPORTANT FIELD SAFETY NOTICE on 09/15/2015. The letter identified the affected device, as well as the reason for the recall, and clinical impact. Customers were informed that both issues would be resolved in Monaco Patch Release 5.10.01. The attached acknowledgement form should be completed, signed, and returned to a local Elekta Office or representative within 30 days. Questions regarding the notice should be directed to a local Elekta office.

Root Cause

Device Design

Distribution Pattern

Worldwide Distribution -- US, including the states of CA, IN, MI, MO, TX and WI; and, the countries of Australia, Canada, Germany, Greece, India, Netherlands, New Zealand, Turkey, and the UK.