Disclaimer: This site aggregates publicly available data from official government sources (FDA, ClinicalTrials.gov, PubMed, SEC EDGAR) for general reference only. It does NOT constitute medical advice, diagnosis, treatment recommendations, or investment advice.

Terminated

K 520411-2, Hercep Test, 35Test, HER2, Protein

FDA Recall Z-0113-06 · Posted 2005-11-01

Recall StatusTerminated
Product CodeMVC
Date Initiated2005-09-29
Date Terminated2007-01-29
Product Quantity545

Reason for Recall

Formulation error in the Visualization Reagent (vial 3) component contained in Hercep Test kits. The error was in the concentration of Polyethylene Glycol contained in the formulation.

Firm Action

Notification letter dated Sept 29,2005 was sent by certified mail to consignees instructing disposal of product.

Root Cause

Other

Distribution Pattern

Nationwide

Official Source

View FDA Recall →