Terminated

Flower Drill Bit Kit 2.7mm- The Drill Bit Kit, 2.7mm (DBK 027) is intended to be used with solid 2.7mm screws and can be used with any the Flower Small and Medium Implant Set Flower Ankle Plating Set , Flower Rear Foot/Calcaneus Plating Set or Flower Upper Extremity Plating Set and uses the medium locking mechanism. Catalog Number: DBK 027

FDA Recall Z-0113-2022 · Posted None

Recall StatusTerminated
Product CodeHTW
Date Initiated2021-09-01
Date Terminated2023-02-16
Product Quantity3 units

Reason for Recall

Drill bits packaged in kits are missing the AO connect feature and may result in delay of procedure

Firm Action

Coventus Orthopedic issued email communication on 9/01/21. Letter states reason for recall, health risk and action to take: Check your inventory for the product listed below. Next, please complete and return the attached Acknowledgement and Receipt Form by Friday, September 10, 2021. If you indicate on the form that you are in possession of affected product, Conventus Flower will ask that you return that affected kits. Conventus Flower will send replacement product as soon as we receive your completed Acknowledgement and Receipt Form. New DBK 020, DBK 024, DBK 027, or LSK 027 kits will be shipped to customers to replace affected product once the Acknowledgement and Receipt Form is received. Other Information: "Contact information for questions: Monday through Friday, 8:30 AM to 5:00 PM, EDST 877-778-8587 or [email protected]

Root Cause

Packaging change control

Distribution Pattern

Arizona California Colorado Florida Florida Illinois Indiana Nevada Ohio Texas Virginia