3M Bair Hugger(TM) Normothermia System, Temperature Monitoring System Sensors Model 360 (Part Number 36000)
FDA Recall Z-0114-2018 · Posted None
Advertisement
Recalling Firm3M Company - Health Care Business
Recall StatusTerminated
Product CodeFLL
Date Initiated2017-10-17
Date Terminated2020-11-03
Product Quantity153 cases (3825 sensors)
Reason for Recall
During a recent investigation, 3M confirmed that a programming translation error could occur in a small amount of the sensors that could lead to a temperature readout that is lower than the patients actual temperature.
Firm Action
The firm initiated their recall on 10/17/2017 via mail. International subsidiaries were notified via email, also on October 17, 2017.Distributors were asked to notify their customers and all direct consignees were directed to dispose of the product on site.
Root Cause
Software design
Distribution Pattern
CA, MI, IL, MO, MN, GA, and Canada, Switzerland
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.