Terminated

Aestiva MRI, Model Numbers: a) 1006-9310-000 b) 1006-9110-000 c) 1006-9023-000 d) 1006-9028-000 e) 1006-9310-000-305077 f) 1006-9310-000-015243 g) 1006-9310-000-017602 h) 1006-9310-000-103785 I) 1006-9310-000-025109 j) 1006-9310-000-009650 k) 1006-9310-000-015224 l) 1006-9310-000-031881 m) 1006-9310-000-031854 n) 1006-9310-000-026571

FDA Recall Z-0114-2020 · Posted None

Recalling FirmGE Healthcare, LLC
Recall StatusTerminated
Product CodeBSZ
Date Initiated2019-07-19
Date Terminated2026-04-13
Product Quantity502 devices

Reason for Recall

Certain Aespire and Aestiva Anesthesia Systems were noted to have a vulnerability to a cyber-attack when connected to the hospital network. An insufficiently secured terminal server may provide an opportunity for a malicious actor that has already penetrated the hospital network to send fraudulent flow sensor correction parameters.

Firm Action

The firm sent an urgent medical device correction notice to their consignees by letter on 07/19/2019. The letter explained the issue and requested that the consignee ensure that the devices are connected to a sufficiently secure terminal server.

Root Cause

Software in the Use Environment

Distribution Pattern

Worldwide