Radiance Data Management System, Software Version 2.60, modification for the ABL800 Flex Blood Gas Analyzer

FDA Recall Z-0115-2007 · Posted 2006-10-28

Recall StatusTerminated
Product CodeJQP
Date Initiated2006-07-06
Date Terminated2011-11-18
Product Quantity5 units

Reason for Recall

Incorrect FI02 values: programming issue can cause incorrect values to be transmitted to the LIS when : 1)- The FIO2 result is edited in the manual sample processing mode--2)- An existing result is opened and FI02 is then edited and sent.....In both cases RADIANCE will transmit the original FI02 value, not the value that was edited.

Firm Action

The recalling firm initially contacted their customers and informed them of the software problem and the upcoming field correction for the software, via telephone on 7/6/2006. A follow-up letter dated 7/21/2006, was also sent to the customers.

Root Cause

Other

Distribution Pattern

Worldwide-The six USA units (equpped with Radiance software) were distributed to a total of two (2) U.S. customers: University of Missouri Hospitals and Clinics in Columbia, MO (4 units) and Vail Valley Medical Center, Vail, CO (2 units). World wide accounts pending submission. Three (3) Canadian customers. They are: 1) Seven Oaks General Hospital MB Canada; and 2) St. Mary''s General Hospital, 911 Queens Blvd, Ontario Canada; and 3) University of Alberta Hospital, 84440 112th Street, Alberta, Canada.