Terminated

MRIdian ViewRay Radiation Therapy System, ViewRay Is indicated for stereotactic radiosurgery and precision radiotherapy for lesions, tumors, and conditions anywhere in the body when radiation treatment is indicated.

FDA Recall Z-0115-2016 · Posted 2015-10-09

Recalling FirmViewray Incorporated
Recall StatusTerminated
Product CodeIYE
Date Initiated2015-08-27
Date Terminated2017-06-09
Product Quantity4 units

Reason for Recall

ViewRay discovered that in the event that an encoder breaks or fails on the Patient Handling System (PHS, or couch), when attempting to restart the system, the couch could move unexpectedly.

Firm Action

ViewRay sent an Customer Advisory Notice dated August 27, 2015, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. It is recommended that if you have a patient on the couch and you receive the fault ''The patient table has experienced an unintended movement. This may be due to the patient or operator leaning on the table. Clear the fault and retry. If the problem persists call Service" on the TPDS,first attempt to clear the fault. If it does not clear and requires you to restart the couch software (same switch as restarting the RTCS) you should manually remove the patient from the couch before restarting the system. A software patch to address this issue is planned for early 2016. In the meantime please practice our recommendation of removing the patient from the couch prior to restarting the RTCS. If you have any questions or concerns regarding this issue, please do not hesitate to contact ViewRay Customer Support at [email protected] or call the ViewRay Customer Support center at 855-286-8875.

Root Cause

Device Design

Distribution Pattern

US Distribution to the states of : CA, MO and WI., and Internationally to Korea.

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.

Official Source

View FDA Recall →