QUIK-COMBO pacing/defibrillation/ECG (therapy) cables used with the LIFEPAK 12 defibrillator/monitor series.
FDA Recall Z-0116-06 · Posted 2005-11-02
Advertisement
Recalling FirmMedtronic Emergency Response Systems, Inc.
Recall StatusTerminated
Product CodeLDD
Date Initiated2005-09-02
Date Terminated2011-01-17
Product Quantitytotal for both types - 52,654
Reason for Recall
Potential for voltage pins in connector of therapy cable to break.
Firm Action
On 9/2/05 the firm issued letters dated September 2005 to their customers. The letters state the problem, advise the cables and/or connectors will be replaced, and emphasize daily inspection and testing,
Root Cause
Other
Distribution Pattern
Devices were distributed worldwide.