Disclaimer: This site aggregates publicly available data from official government sources (FDA, ClinicalTrials.gov, PubMed, SEC EDGAR) for general reference only. It does NOT constitute medical advice, diagnosis, treatment recommendations, or investment advice.

Terminated

Channel Steerable Sheath in 8F Enpath part # 10775-003, BARD Part Number XD10775003 (US Distribution Only) and 9F Enpath part # 10775-004 (foreign distribution), BARD Part Number XD10775004; Labeled as Distributed by Bard Electrophysiology Division.

FDA Recall Z-0118-2007 · Posted 2006-11-02

Recalling FirmEnpath Medical, Inc.
Recall StatusTerminated
Product CodeDQY
Date Initiated2006-06-27
Date Terminated2007-04-14
Product Quantity2494 Units

Reason for Recall

Enpath has become aware that some of the Enpath Medical 8F & 9F Steerable sheath devices may have a non-conforming flushport bond that causes a leak where air may enter the flush port chamber.

Firm Action

Enpath''s three Distributors were notified via letter or telephone on June 27th, 29th and July 6th with a description of the problem and further distribution or use of any remaining product should cease immediately. Instructed to notify their customers about the product and forward to Enpath for evaluation.

Root Cause

Other

Distribution Pattern

CA, MA

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.

Official Source

View FDA Recall →

Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.

↑