Terminated

AS-OBGYN Information System version 7.824.x

FDA Recall Z-0118-2019 · Posted None

Recalling FirmAS Software, Inc.
Recall StatusTerminated
Product CodeLLZ
Date Initiated2018-09-20
Date Terminated2019-04-19
Product Quantity52 systems

Reason for Recall

Any change made to current ultrasound machine software or hardware upgrades, adding or modifying custom fields, or adding a new Ultrasound Machine, must be tested prior to the use in production environment to ensure that ultrasound measurements are crossing over correctly to the AS system

Firm Action

AS Software issued letter by email on 9/20/18 to notify customers that any change made to their ultrasound machine such as software or hardware upgrades, adding or modifying custom fields, or adding a new Ultrasound Machine, are tested prior to the use in the production environment. Medical Device Correction Return Response form to be signed and returned. The AS Software Support Department can be reached Monday through Friday, 9:00 AM to 6:00 PM EST via phone: 201-541-1900 option 2 or e-mail: [email protected]

Root Cause

Device Design

Distribution Pattern

US Nationwide.