Terminated

Michigan Instruments Thumper Mechanical CPR Device; Model 1007.

FDA Recall Z-0119-2007 · Posted 2006-11-03

Recall StatusTerminated
Product CodeDRM
Date Initiated2006-08-04
Date Terminated2006-12-06
Product Quantity1100

Reason for Recall

Failure to initiate compressions when first turned on, if improperly shut down-Operator Manual updated to include proper shut down procedures

Firm Action

Consignees were sent a letter dated 8/4/06 with additional instructions for use. A second letter was sent on 9/1/06 to correct the new instructions provided in the first letter for page 19 of the instructions.

Root Cause

Other

Distribution Pattern

Worldwide- USA and countries of Brazil, Canada, China, Germany, Greece, Japan, Korea, Macau, Netherlands, Pakistan, Spain, Taiwan and United Kingdom.

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.

Official Source

View FDA Recall →