MEDLINE ReNewal Ligasure Sealer/Divider Nano-Coated Compatible with FT10 Generator, REF LF4418CR. electrosurgical instrument

FDA Recall Z-0119-2024 · Posted 2023-10-16

Recall StatusOpen, Classified
Product CodeNUJ
Date Initiated2023-08-03
Product Quantity268 units

Reason for Recall

Various reprocessed Covidien LigaSure and Ethicon Harmonic devices may have an incomplete seal on the packaging tray.

Firm Action

Medline issued a recall notice to its consignees by email on 08/03/2023. The notice explained the problem and the potential risk. Those consignees who have transferred possession of the affected product were directed to notify their customers. The firm is seeking the return of the affected product. An error in the model number was noted (missing an "R" at the end) and the firm issued an updated letter on 09/19/2023 with the corrected number.

Root Cause

Packaging

Distribution Pattern

US